ELP

ELP-01D - Electroporator for Tissue and Cultured Cells

Electroporation is one method to transfect multiple cells in cell culture or tissue.

The ELP-01D is an amplifier providing square wave voltage pulses of up to 110 V which can be comfortable timed with the internal digital timing unit. An electrode resistance test circuit completes the functionalilty of this instrument.

Features

  • Square wave pulse generation, isolated output
  • Single, multiple (up to 999) pulses, or continuous cycling operation
  • Digital cycle counter (000 to 999 cycles)
  • STOP and RESET function (switch operated)
  • DELAY and PULSE duration up to 9999.9 ms; resolution 0.1 ms
  • Manual, triggered (TRIGGER TTL) or gated (INPUT TTL) operation
  • AMPLITUDE setting with digital potentiometer (±110 V, 1 V resolution)
  • Electrode resistance test (up to 9.99 kΩ, resolution 10Ω)

Ordering

Part No. Description
ELP-01D Electroporator/Transfector for Tissue and Cell Cultures

Technical Data

PULSE (x 0. 1 ms) switch:
5 digits thumb wheel, 0.1 ms to 9999.9 ms, digital

DELAY (x 0. 1 ms) switch:
5 digits thumb wheel, 0.1 ms to 9999.9 ms, digital

PULSE AMPLITUDE (x1V) switch:
3 digits thumb wheel, 0 V to 110 V, digital

TRIGGER IN:

TTL signal, 10 µs minimum

ISOLATED OUTPUT:
±110 V max., isolated, set in 1 V steps

CURRENT OUTPUT:

BNC connector, 1 V / mA

POWER SUPPLY:
external, 13.8 V, DC