ISO-STIM-01M

ISO-STIM 01M - Isolated Stimulator

Features:

  • NO batteries needed
  • ISOLATED OUTPUT
  • Three input modes: ANALOG, GATE TTL, DIRECT
  • Stimulus AMPLITUDE set by three digit control and a RANGE switch (CURRENT) or by an external source, e.g. a data acquisition system
  • Stimulus duration set by two digit control from an internal, digital timer (10 ìs to 99 ms) or by an external source
  • Stimulus polarity set by switch (+, 0, -) or by data acquisition system
  • Current (±10 mA) or voltage (±100 V) source output
  • “Overrange” and “active” LED indicators
  • Capacitance compensation for the electrode capacitance

Combine with

  • EXT-10C (extracellular amplifier module)
  • DPA 2FX (amplifier / filter module)
  • BA-01M (intracellular amplifier)

to get a complete system for stimulating, extracellular and intracellular recording.

Also available as stand-alone device in small desktop housing.

Ordering

Part no. Description
ISO-STIM 01M-100 Stimulus isolation module (+/-100V, +/- 10 mA), with bipolar stimulus output, width = 2
ISO-CAB-M Cable set for connecting the stimulation electrode

Technical Data

INPUT OPERATING MODES:
DIRECT, ANALOG TRIGGER, TTL TRIGGER, selectable with toggle switch

FUNCTION MODES:
voltage source, current source, selectable with toggle switch

SOURCE RANGES:
±10 mA (current source),
±100 V (voltage source), selectable with toggle switch

TIME RANGE FACTORS:

x10 µs, x100 µs, x1ms, selectable with toggle switch, time range up to 99 ms

AMPLITUDE POTENTIOMETER:

3 digits, ±10 V or ±10 mA max.,
polarity selectable with toggle switch

CURRENT RANGE FACTORS:

x0.1 µA, x1 µA, x10 µA,
selectable with toggle switch

TIME POTENTIOMETER:

2 digits, 0 - 99 as multiplication factor selectable

INPUT VOLTAGE RANGE:

±10 V, amplified internally with gain of 10 (voltage source) or gain of 1 (current source)

OUTPUT VOLTAGE:
±100 V max., isolated

OUTPUT CURRENT:

±10 mA max., isolated